A method for accurate zero calibration of asymmetric jaws in single-isocenter half-beam techniques

Autor
Hernandez, V.; Sempau, J.; Abella, R.; Lopez, M.; Perez, M.; Artigues, M.; Arenas, M.
Tipus d'activitat
Article en revista
Revista
Medical physics
Data de publicació
2013-02
Volum
40
Número
2
Pàgina inicial
1
Pàgina final
6
DOI
https://doi.org/10.1118/1.4773314 Obrir en finestra nova
Repositori
http://hdl.handle.net/2117/18727 Obrir en finestra nova
URL
http://www.scopus.com/record/display.url?eid=2-s2.0-84873586272&origin=inward&txGid=D24C552D0A97633B3A6ECBFF4FC06A45.WlW7NKKC52nnQNxjqAQrlA%3a7 Obrir en finestra nova
Resum
Purpose: To present a practical method for calibrating the zero position of asymmetric jaws that provides higher accuracy at the central axis and improves dose homogeneity in the abutting region of half-beams. Methods: Junction doses were measured for each asymmetric jaw using the double-exposure technique and electronic portal imaging devices. The junction dose was determined as a function of jaw position. The shift in the zero jaw position (or in its corresponding potentiometer readout) requir...
Citació
Hernandez, V. [et al.]. A method for accurate zero calibration of asymmetric jaws in single-isocenter half-beam techniques. "Medical physics", Febrer 2013, vol. 40, núm. 2, p. 1-6.
Grup de recerca
CREB - Centre de Recerca en Enginyeria Biomedica
DRM - Dosimetria i Radiofísica Mèdica

Participants

  • Hernandez Espinosa, Victor  (autor)
  • Sempau Roma, Josep  (autor)
  • Abella, R.  (autor)
  • Romero Lopez, Matilde  (autor)
  • Lopez Perez, Mercedes  (autor)
  • Artigues, Manel  (autor)
  • Arenas, M.  (autor)