Carregant...
Carregant...

Vés al contingut (premeu Retorn)

Towards the standardization of ballistocardiography systems for J-peak timing measurement

Autor
Gomez-Clapers, J.; Serra, A.; Casanella, R.; Pallas-Areny, R.
Tipus d'activitat
Article en revista
Revista
Measurement
Data de publicació
2014-12-01
Volum
58
Pàgina inicial
310
Pàgina final
316
DOI
https://doi.org/10.1016/j.measurement.2014.09.003 Obrir en finestra nova
Projecte finançador
INSTRUMENTACION NO INVASIVA PARA MONITORIZAR Y EVALUAR TENDENCIAS EN EL SISTEMA CARDIOVASCULAR
Repositori
http://hdl.handle.net/2117/24378 Obrir en finestra nova
URL
http://www.sciencedirect.com/science/article/pii/S0263224114003820 Obrir en finestra nova
Resum
There is a growing interest in accurately measuring the timing of the J peak of the ballistocardiogram (BCG) in order to obtain cardiovascular function markers non-invasively, especially in modern home healthcare applications. In this paper we have studied the effect that some common uncertainty sources have in the time measurement of the J peak. This is a necessary step towards the standardization of modern ballistocardiography systems equivalent to that available for ECG systems. We conclude t...
Citació
Gomez-Clapers, J. [et al.]. Towards the standardization of ballistocardiography systems for J-peak timing measurement. "Measurement", 16 Setembre 2014, vol. 58, p. 310-316.
Paraules clau
Acquisition system, Ballistocardiogram, Biomedical instrumentation, Standardization, WEIGHING SCALE
Grup de recerca
GRUP ISI - Grup d'Instrumentació, Sensors i Interfícies

Participants