Carregant...
Carregant...

Vés al contingut (premeu Retorn)

Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants

Autor
Gil, F.J.; Herrero-Climent, M.; Lázaro, P.; Rios, J.
Tipus d'activitat
Article en revista
Revista
Journal of materials science. Materials in medicine
Data de publicació
2014-04-10
Volum
25
Número
7
Pàgina inicial
1825
Pàgina final
1830
DOI
https://doi.org/10.1007/s10856-014-5211-7 Obrir en finestra nova
Repositori
http://hdl.handle.net/2117/27711 Obrir en finestra nova
Resum
Microgap between implant and abutment can produce biological and mechanical problems such as peri- implantitis and/or fatigue failures. The aim of this study was to evaluate microgap size and fatigue behavior of external and internal connections. In both systems the torque to tighten the abutment screw of single crown abutments was 45 Ncm. Fifty implants for each connection type were studied. One subgroup ( n = 5) was used by the observation and evaluation of ...
Citació
Gil, F.J. [et al.]. Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants. "Journal of materials science. Materials in medicine", 10 Abril 2014, vol. 25, núm. 7, p. 1825-1830.
Grup de recerca
BBT - Biomaterials, Biomecànica i Enginyeria de Teixits
CREB - Centre de Recerca en Enginyeria Biomedica
CRnE - Centre de Recerca en Ciència i Enginyeria Multiescala de Barcelona

Participants

  • Gil Mur, Francisco Javier  (autor)
  • Herrero Climent, Mariano  (autor)
  • Lázaro, Pedro  (autor)
  • Rios Santos, Jose Vicente  (autor)