A novel method to obtain proximal plethysmographic information from distal measurements using the impedance plethysmogram

Autor
Gomez-Clapers, J.; Casanella, R.; Pallas-Areny, R.
Tipus d'activitat
Article en revista
Revista
Journal of Electrical Bioimpedance
Data de publicació
2015-12-23
Volum
6
Pàgina inicial
44
Pàgina final
48
DOI
https://doi.org/10.5617/jeb.2575 Obrir en finestra nova
Projecte finançador
EEBB-I-13-06744
INSTRUMENTACION NO INVASIVA PARA MONITORIZAR Y EVALUAR TENDENCIAS EN EL SISTEMA CARDIOVASCULAR
Repositori
http://hdl.handle.net/2117/87689 Obrir en finestra nova
URL
https://www.journals.uio.no/index.php/bioimpedance/article/view/2575 Obrir en finestra nova
Resum
We present a novel method to detect proximal volume changes based on the impedance plethysmogram (IPG) measured from limb to limb with two electrode pairs symmetrically placed at distal areas of the upper or the lower limbs. Since the measurement is sensitive to changes along the whole current path, this method allows us to detect changes in arteries that are more proximal to the torso than the measurement sites. Our results show that the Pulse Arrival Time (PAT) measured from the R peak of the ...
Citació
Gomez-Clapers, J., Casanella, R., Pallas, R. A novel method to obtain proximal plethysmographic information from distal measurements using the impedance plethysmogram. "Journal of Electrical Bioimpedance", 23 Desembre 2015, vol. 6, p. 44-48.
Paraules clau
IPG, PTT, cardiovascular monitoring, limbs