On the feasibility of noncontact ECG measurements

Autor
Pallas-Areny, R.; Casas, J.O.
Tipus d'activitat
Article en revista
Revista
IEEE transactions on instrumentation and measurement
Data de publicació
2016-03-01
Volum
65
Número
3
Pàgina inicial
727
Pàgina final
728
DOI
https://doi.org/10.1109/TIM.2016.2516778 Obrir en finestra nova
Repositori
http://hdl.handle.net/2117/102390 Obrir en finestra nova
URL
http://ieeexplore.ieee.org.recursos.biblioteca.upc.edu/stamp/stamp.jsp?arnumber=7401243 Obrir en finestra nova
Resum
The article by Kranjec et al. [1], “Novel methods of noncontact heart rate measurement: A feasibility study” is interesting and informative as it compares different contactless methods for heart rate detection. Nevertheless, the use of the term “capacitively coupled ECG” (CCECG) in the article is confusing and may mislead readers. That article studies the feasibility of four noncontact methods for heart rate measurement, which are classified in two groups: “the methods measuring electr...
Citació
Pallas-Areny, R., Casas, J. On the feasibility of noncontact ECG measurements. "IEEE transactions on instrumentation and measurement", 1 Març 2016, vol. 65, núm. 3, p. 727-728.
Paraules clau
Biomedical measurement, Capacitance, Capacitance measurement, Electrocardiography, Electrodes, Heart rate measurement, Impedance, Surface impedance