A practical approach to perform the isotropy test for extremity dosemeters

Autor
Gultresa, J.; Llansana, J.; Roig, M.; Ginjaume, M.
Tipus d'activitat
Article en revista
Revista
Radiation protection dosimetry
Data de publicació
2016-09-01
Volum
170
Número
1-4
Pàgina inicial
95
Pàgina final
99
DOI
https://doi.org/10.1093/rpd/ncv530 Obrir en finestra nova
URL
https://academic.oup.com/rpd/article-abstract/170/1-4/95/2223851/A-PRACTICAL-APPROACH-TO-PERFORM-THE-ISOTROPY-TEST?redirectedFrom=fulltext Obrir en finestra nova
Resum
The requirements for determining extremity dosemeter performance have evolved over the past decade. In 2010, the Spanish Nuclear Safety Council (CSN) adopted a national protocol with performance requirements based on ISO 12794. Because of the lack of symmetry of ISO 4037-3 calibration phantoms, the isotropy test set up in the sagittal plane presented several challenges both for individual monitoring services and for calibration labs. This article proposes and validates a practical approach to re...