Air kermato H-P(3) conversion coefficients for IEC 61267 RQR X-RAY radiation qualities: application to dose monitoring of the lens of the eye in medical diagnostics

Autor
Principi, S.; Guardiola, C.; Duch, M.; Ginjaume, M.
Tipus d'activitat
Article en revista
Revista
Radiation protection dosimetry
Data de publicació
2016-09-01
Volum
170
Número
1-4
Pàgina inicial
45
Pàgina final
48
DOI
https://doi.org/10.1093/rpd/ncv435 Obrir en finestra nova
URL
https://academic.oup.com/rpd/article-abstract/170/1-4/45/2463408/AIR-KERMA-TO-Hp-3-CONVERSION-COEFFICIENTS-FOR-IEC?redirectedFrom=fulltext Obrir en finestra nova
Resum
Recent studies highlight the fact that the new eye lens dose limit can be exceeded in interventional radiology procedures and that eye lens monitoring could be required for these workers. The recommended operational quantity for monitoring of eye lens exposure is the personal dose equivalent at 3 mm depth Hp(3) (ICRU 51). However, there are no available conversion coefficients in international standards, while in the literature coefficients have only been calculated for monoenergetic beams and f...