Effect of oblique X-ray incidence in flat-panel computed tomography of the Breast

Autor
Badano, A.; Kyprianou, I.; Freed, M.; Jennings, R.; Sempau, J.
Tipus d'activitat
Article en revista
Revista
IEEE transactions on medical imaging
Data de publicació
2009
Volum
28
Número
5
Pàgina inicial
696
Pàgina final
702
Repositori
http://hdl.handle.net/2117/9572 Obrir en finestra nova
Resum
We quantify the variation in resolution due to anisotropy caused by oblique X-ray incidence in indirect flat-panel detectors for computed tomography breast imaging systems. We consider a geometry and detector type utilized in breast computed tomography (CT) systems currently being developed. Our methods rely on MANTIS, a combined X-ray, electron, and optical Monte Carlo transport open source code. The physics models are the most accurate available in general-purpose Monte Carlo packages in the d...
Citació
Badano, A. [et al.]. Effect of oblique X-ray incidence in flat-panel computed tomography of the Breast. "IEEE transactions on medical imaging", 2009, vol. 28, núm. 5, p. 696-702.
Grup de recerca
CREB - Centre de Recerca en Enginyeria Biomedica
DRM - Dosimetria i Radiofísica Mèdica