Direct pulse transit time measurement from an electronic weighing scale

Author
Gomez-Clapers, J.; Casanella, R.; Pallàs, R.
Type of activity
Presentation of work at congresses
Name of edition
Computing in Cardiology 2016
Date of publication
2016
Presentation's date
2016-09-14
Book of congress proceedings
Computing in Cardiology 2016, Vol. 43
First page
773
Last page
776
Repository
http://hdl.handle.net/2117/102901 Open in new window
URL
http://ieeexplore.ieee.org/document/7868857/ Open in new window
Abstract
A novel method to measure Pulse Transit Time (PTT) from the force sensors of a modified electronic weighing scale is presented. The method relies on the relation between PTT and the interval between I and J waves of the ballistocardiogram (BCG). The IJ interval and the PTT derived from a carotid and a femoral pressure-pulse waveform were simultaneously obtained from three subjects during paced respiration, which induces periodic hemodynamic changes. The correlation between the IJ interval and th...
Citation
Gomez-Clapers, J., Casanella, R., Pallas-Areny, R. Direct pulse transit time measurement from an electronic weighing scale. A: Computing in Cardiology Annual Conference. "Computing in Cardiology 2016, Vol. 43". Vancouver: 2016, p. 773-776.
Keywords
Correlation, Electrocardiography, Electrodes, Hypertension, Pulse measurements, Sensors, Velocity measurement