Loading...
Loading...

Go to the content (press return)

Understanding electrosurgical unit perturbations in order to address hospital operating room electromagnetic compatibility

Author
Fernandez, M.; Quilez, M.; Silva, F.
Type of activity
Journal article
Journal
IEEE transactions on biomedical engineering
Date of publication
2006-06
Volume
53
Number
6
First page
1026
Last page
1209
URL
http://hdl.handle.net/2117/1346 Open in new window
Abstract
We analyze the radiated emissions from a low power electrosurgical unit (ESU). Measurements at a 3-meter distance are performed in order to find a suitable measurement setup intended to reproduce realistically the real working ESU behavior in a test lab. The broad-band and narrow-band characteristics of the perturbation are studied in order to address the other equipment immunity.
Citation
Fernandez Chimeno, M; Quilez, M; Silva, F. Understanding electrosurgical unit perturbations in order to address hospital operating room electromagnetic compatibility. IEEE Transactions on Biomedical Engineering, 2006 vol.53, núm. 6 p.1206-1209
Group of research
CREB - Biomedical Engineering Research Centre
GCEM - Electromagnetic Compatibility Group
GRUP ISI - Instrumentation, Sensors and Interfaces Group
IEB - Electronic and Biomedical Instrumentation

Participants

Attachments