Error analysis in two-terminal impedance measurements with residual correction

Author
Torrents, J.M.; Pallàs, R.
Type of activity
Journal article
Journal
IEEE transactions on instrumentation and measurement
Date of publication
2005-10
Volume
54
Number
5
First page
2113
Last page
2116
URL
http://hdl.handle.net/2117/1388 Open in new window
Abstract
Residual impedance correction in impedance analyzers when using an asymmetrical test fixture needs three reference measurements, usually open circuit, short circuit, and load (meaning an impedance close to the impedance under test). This paper provides an error estimate for impedance measurements that apply a simple open/short correction in spite of using an asymmetrical test fixture. Experimental results show that the minimal error is obtained for impedance values close to the geometric mean of...
Citation
Torrents, J.M.; Pallas R. Error analysis in two-terminal impedance measurements with residual correction. IEEE Transactions on Instrumentation and Measurement. 2005, vol 54 núm 5, p. 2113-2116